Enviro-LOGO

Thickness Meter PCE-CT 80

Category Tag

The paint layer thickness gauge PCE-CT 80 is a measuring device for the nondestructive measurement of coatings (paints, lacquers, plastics …) on steel / iron and non-ferrous metals. Thanks to the externally connected sensor on the PCE-CT 80 ink layer thickness gauge, even difficult-to-reach measuring points can be easily achieved. The menu navigation of the paint thickness gauge allows easy adjustment and setting to new parameters and makes this handy paint coating thickness gauge an indispensable tool for control measurements in production, workshop and quality assurance. Includes FN1.5 probe in kit for 0 … 1500 μm or 0 … 59 mils measurements.

Measuring range Fe: 0 to 5000 μm / 0 … 196.9 mils (depending on sensor)
NFe: 0 to 3000 μm / 0 … 118.1 mils (depending on sensor)
Accuracy ± (2% of reading + 1 μm / 0.039 mils)
Resolution 0.1 μm (0.0039 mils) (< 100 μm / 3.94 mils)
1 μm / 0.039 mils (> 100 μm / 3.94 mils)
Measurable materials Non-magnetic layers on steel, iron, …
Non-electrically conductive layers on aluminum, copper, …
Min. radius of curvature convex 5 mm / 0.2 in
Min. radius of curvature      concave 25 mm / 1 in
Min. measuring surface Ø17 mm / 0.7 in
Min. substrate thickness 0.2 mm / 0.08 in (on magnetic materials)
0.05 mm / 0.002 in (on non-magnetic materials)
Probe mode Autom. Mode with material detection (Fe + NFe)
Magnetic mode (Fe)
Eddy current mode (NFe)
Measurement mode Single measurement
continuous measurement
Calibration Multipoint calibration (1 … 4 points for each group)
Zeroing
Units μm, mm, mils
Data transfer USB 2.0
Storage A volatile measuring group (DIR mode)
Four measuring groups with autom. Storage and max. 2000 readings (GEN mode)
Statistical functions Number of Measured values, mean, minimum, maximum, standard deviation
Alarm Display when the adjustable upper and lower alarm limits are exceeded
Operating time Autom. Shutdown mode (3 min.)
Power supply 3 x 1.5V AAA batteries
Display 128 x 128 px LCD display
Indicators Battery status
Fault detection
Operating conditions 0 … 50°C / 32 … 122°F
20 … 90% RH not condensing
Storage conditions -10 … 60°C / 14 … 140°F
20 … 90% RH not condensing
Dimensions (L x W x H) 143 x 71 x 37 mm / 5.6 x 2.8 x 1.5 in
Weight With sensor and batteries: approx. 271 g / < 1 lb