testo 330-2 LL

SKU 0632 3307 70 Categories , Tag

testo 330-2 LL flue gas analyzer with long-life gas sensors and built-in draugh and gas zeroing; BLUETOOTH and with H2-compensated CO-sensor; incl. rech. battery and calibration protocol; with graphic display; version 2010

Details

Product Description
Thanks to the new instrument functions, the Testo flue gas analyzers of the new testo 330-LL series now offer you even more professional and reliable support. These are the main features:
β€’ High-resolution colour display for the graphic representation of your measurement data
β€’ Extended measurement menus – such as gas pipe testing – for a comprehensive analysis of your heating system
β€’ Logger function for easy long-term recording of the measurement curve

Technical Data

General technical data
Weight 600 g (without rechargeable battery)
Dimensions 270 x 90 x 65 mm
Operating temperature -5 to +45 Β°C
Display size 240 x 320 pixels
Display function Colour graphic display
Power supply Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A
Maximum memory 500,000 readings
Storage temperature -20 to +50 Β°C
Differential Pressure – Piezoresistive
Measuring range Β±10000 Pa
Accuracy Β±0.3 Pa (0 to 9.99 Pa) plus Β±1 Digit

Β±3 % of mv (10 to 10000 Pa) plus Β±1 Digit

Flue gas Oβ‚‚
Measuring range 0 to 21 Vol.%
Accuracy Β±0.2 Vol.%
Resolution 0.1 Vol.%
Reaction time t₉₀ < 20 s
Flue gas CO (with Hβ‚‚-compensation)
Measuring range 0 to 8000 ppm
Accuracy Β±10 ppm or Β±10 % of mv (0 to 200 ppm)

Β±20 ppm or Β±5 % of mv (201 to 2000 ppm)

Β±10 % of mv (2001 to 8000 ppm)

Resolution 1 ppm
Reaction time t₉₀ < 60 s
Flue gas COlow (with Hβ‚‚-compensation)
Measuring range 0 to 500 ppm
Accuracy Β±2 ppm (0 to 39.9 ppm)

Β±5 % of mv (40 to 500 ppm)

Resolution 0.1 ppm
Reaction time t₉₀ < 40 s
CO determination (with Hβ‚‚-compensation), automatic dilution
Measuring range 0 to 30000 ppm
Accuracy Β±100 ppm (0 to 1000 ppm)

Β±10 % of mv (1001 to 30000 ppm)

Resolution 1 ppm
Flue gas NO
Measuring range 0 to 3000 ppm
Accuracy Β±5 ppm (0 to 100 ppm)

Β±5 % of mv (101 to 2000 ppm)

Β±10 % of mv (2001 to 3000 ppm)

Resolution 1 ppm
Reaction time t₉₀ < 30 s
Flue gas Nolow
Measuring range 0 to 300 ppm
Accuracy Β±2 ppm (0 to 39.9 ppm)

Β±5 % of mv (40 to 300 ppm)

Resolution 0.1 ppm
Reaction time t₉₀ < 30 s
Differential Pressure flue gas Draught – Piezoresistive
Measuring range -9.99 to +40 hPa
Accuracy Β±0.02 hPa or Β±5 % of mv (-0.50 to +0.60 hPa)

Β±0.03 hPa (+0.61 to +3.00 hPa)

Β±1.5 % of mv (+3.01 to +40.00 hPa)

Resolution 0.01 hPa
Temperature
Measuring range -40 to +1200 Β°C
Accuracy Β±0.5 Β°C (0 to +100.0 Β°C)

Β±0.5 % of mv (Remaining Range)

Resolution 0.1 Β°C (-40 to +999.9 Β°C)

1 Β°C (> +1000 Β°C)

Flue gas degree of effectivity, Eta (calculated)
Measuring range 0 to 120 %
Resolution 0.1 %
Flue gas loss (calculated)
Measuring range 0 to 99.9 %
Resolution 0.1 %
Flue gas COβ‚‚ calculation (calculated from Oβ‚‚)
Measuring range 0 to COβ‚‚ max (Display range)
Accuracy Β±0.2 Vol.%
Resolution 0.1 Vol.%
Reaction time t₉₀ < 40 s
Velocity / Volume flow
Measuring range 0.15 to 3 m/s
Resolution 0.1 m/s
Flue gas CO (without Hβ‚‚-compensation)
Measuring range 0 to 4000 ppm
Accuracy Β±20 ppm (0 to 400 ppm)

Β±5 % of mv (401 to 2000 ppm)

Β±10 % of mv (2001 to 4000 ppm)

Resolution 1 ppm
Reaction time t₉₀ < 60 s
Ambient CO
Measuring range 0 to 500 ppm
Accuracy Β±5 ppm (0 to 100 ppm)

Β±5 % of mv (> 100 ppm)

Resolution 1 ppm
Reaction time Approx. 35 s

with CO probe

Ambient COβ‚‚
Measuring range 0 to 1 Vol.%

0 to 10000 ppm

Accuracy Β±50 ppm or Β±2 % of mv (0 to 5000 ppm)

Β±100 ppm or Β±3 % of mv (5001 to 10000 ppm)

Reaction time Approx. 35 s

with ambient COβ‚‚ probe

Gas leak measurement for combustible gases (via gas leak detection probe)
Measuring range 0 to 10000 ppm CHβ‚„/ C₃Hβ‚ˆ; Display range
Accuracy Signal optical display (LED) audible signal via buzzer
Reaction time t₉₀ < 2 s

with gas leak detection probe

Temperature (via the fine pressure probe)
Measuring range -40 to +1200 Β°C max. (dependent on probe)
Accuracy Β±0.5 Β°C (-40 to 100 Β°C)

Β±0.5 % of mv (Remaining Range) plus probe accuracy

Resolution 0.1 Β°C